Product Center
产品中心
TORCH Rapid Tests
TORCH-IgM IgG Rapid test
Catalog No.:BG801C BG802C
INTENDED USE
The TORCH test is a panel of rapid qualitative lateral flow test designed for the qualitative detection of IgG/IgM antibodies to Toxoplasma gondii(TOXO), Cytomegalovirus (CMV), Rubella, Herpes Simplex virus (HSV-2) in human serum/plasma samples.
The TORCH test is a panel of rapid qualitative lateral flow test designed for the qualitative detection of IgG/IgM antibodies to Toxoplasma gondii(TOXO), Cytomegalovirus (CMV), Rubella, Herpes Simplex virus (HSV-2) in human serum/plasma samples.
Test Procedures
1.Dispense 5 ul of specimen into location of allow (between T line region and end edge of the view window). The T line region must be wetted by the sample added.
2.After 30 seconds, add 2 drops of Develop Buffer into each sample well.
3.Read time:
Torch-IgM Panel Test: 10 minutes
Torch -IgG Panel Test: 10 minutes
Interpretation of Results
Positive Result: If there is a rose-pink color band in the control region (marked with a "C"), and a rose-pink color band in the test region (marked with a "T"), TORCH antibody is present and the specimen is positive.
Negative Result: The absence of a color band in the test region next to the letter "T" indicates the absence of any detectable TORCH antibody.
Invalid Result: If a color band does not appear in the control region "C", the test results are invalid. The sample may have been added to the wrong window, or the Test Device may have deteriorated. This specimen should be re- tested using a new Test Device.