Troponin I(cTnl)Rapid Tests

+
  • zaza.png

Troponin I(cTnl) test cassette 
Catalog No.: BGW1401C

INTRODUCTION

Troponin I (cTnl) Test is a qualitative, membrane based immunoassay for the detection of cTnl in whole blood, serum or plasma. The membrane ist pre-coated with streptavidin on the test line region of the test. During testing, the whole blood, serum or plasma specimen reacts with the particle coated with anti-cTnl antibodies and biotinylated anti-cTnl antibodies. The mixture mi grates upward on the membrane
chromatographically by capillary action to react with streptavidin on the membrane and generate a red line. The presence of this red line in the test line region indicates a positive  result, while its absence indicates a negative result. To serve as a procedural control, a red line will always appear in the control line region indicating that proper volume of specimen has been added and membrane wicking has occurred.

TEST PROCEDURE

1. Remove the test card from the pouch and allow it to equilibrate to room temperature prior to testing. Lay the card flat on a clean, dry and non-absorbent surface. 
2. Dispense 2 drops serum specimen (100ul) into the specimen well.
3. Wait 20 minutes and read result. Do not read results after 20 minutes.

INTERPRETATION OF RESULTS

Negative: Only one colored band appears on the control (C) region. No apparent band on the test (T) region. 
Positive: in addition to a pink colored control (C) band, a distinct pink colored band will also appear in the test (T) region. Invalid: A total absence of color in both regions is an indication of procedure error and/or the test reagent has deteriorated. 